亚洲成人电影院av_亚洲欧美日韩男同_一本欧美日韩中文字幕_部长连续几天让我失去理智系列小说_亚洲综合高清无码_韩国一级毛片a级免观看_欧美性猛交xxx乱大交_国产无遮挡又黄又爽的网站_四虎成人在线_人妻丰满熟妇aⅤ无码久久洗澡

DEARTO-Professional manufacturer of temperature and humidity measuring instruments
Tai'an Detu Automation Instrument Co., Ltd.

0538-5089056

Surface Sensor Calibrator - Surface Temperature Probe Calibration System

640480-1.jpg

◇ Introduction ◇ Function◇ Parameter◇ Product order◇ Download◇ Related products

Introduction

Surface Sensor Calibrator - Surface Temperature Probe Calibration System is mainly used to calibrate surface sensors of various sizes and types, such as thermistors, thin film sensors, surface resistance thermometers, ribbon sensors, surface thermocouples, etc. 

DTZ-400 surface sensor calibrator is composed of a controller and a heater. It is easy to operate, intuitive to display, and can be heated simply by setting the temperature. The temperature rises quickly, and the multi-stage PID temperature control is accurate. The effective use range is up to 130mm in diameter, and has a communication interface that can communicate with a computer. 

DTZ-400 surface probe calibrator can be used as a calibration source for thermal imaging cameras with slight changes.

Function

Large calibration plane: the calibration plane has a diameter of 130mm, and can calibrate various types of surface-mount temperature sensors

Rapid temperature rise: from room temperature to 400 ℃ for 30min

High stability: ±0.3℃@300℃, ±0.4℃@400℃

Stabilization time: only 8min to stabilize

High resolution: resolution is 0.01 ℃

Portable and reliable: the whole machine weighs only 3.5KG, easy to carry

Multi-segment temperature control: Built-in multi-segment PID can realize segmented temperature control, small overshoot and overshoot, rapid temperature rise, and small temperature control fluctuation

Communication function: RS-232 standard interface, can read and control the temperature of the calibration system

◆ Calibration temperature can reach 400 ℃

DTZ-400 surface sensor calibrator has a good temperature rise and fall speed, from room temperature to 400 ℃, it takes 30 minutes, from 400 ℃ to room temperature, it takes 55 minutes, with rapid cooling device only 50 minutes, greatly improved work efficiency.

◆ The diameter of the calibration plane reaches 130 mm, which can meet the calibration needs of most surface temperature sensors

DTZ-400 ssurface temperature calibration diameter reaches 130mm, enough to calibrate surface sensors of various sizes and types, such as thermistors, thin film sensors, surface resistance thermometers, ribbon sensors, surface-mount thermocouples, etc.

◆ External reference platinum resistance thermometer can be installed to reduce the uncertainty of calibration

In order to achieve the ultimate calibration accuracy, the design engineer reserved a thermometer socket at the bottom of the aluminum surface. A platinum resistance thermometer with a diameter of 5mm can be inserted as a standard instrument for comparison and calibration, which greatly improves the calibration accuracy.

◆ Excellent temperature uniformity and surface finish of the calibration plane

Surface sensors are often difficult to calibrate because it is difficult to find a flat, uniform surface temperature source. The DTZ-400 surface temperature calibration system uses aluminum material as the surface and has a high finish. It can ensure that the sensor being calibrated has good contact with the surface temperature calibrator, thereby reducing the calibration uncertainty. DTZ-400 uses special heating materials and constant temperature structure, the surface temperature is more uniform.

◆ Accurate and stable temperature control, small overshoot temperature

The device provides a variety of parameter options for users to choose, including setting the PID parameters of three different temperature ranges of high, medium and low, the inflection point between different temperatures, alarm temperature and so on.

◆ Portable and easy to use

The whole machine weighs only 3.5kg, compact and portable.

Technical indicators

無(wú)標(biāo)題文檔

Model

 DTZ-400G surface sensor calibrator

 DTZ-400BG surface sensor calibration

DTZ-400BCG surface sensor calibration

DTZ-450BG surface probe calibrator

DTZ-460BG surface probe calibrator

Temperature range

35℃-400℃

35℃-400℃

50℃~400℃

50~500℃

50~500℃

product type

Regular

Applicable for L-type sensor

Smart touch screen

High temperature customization

Calibration plane diameter

130mm

Uniformity

0.4℃@35~ 100℃ ,0.8℃@100~300℃,1.2℃@300~400℃

Volatility(10min)

0.2℃@35~ 100℃,0.4℃@100~300℃,0.8℃@300~400℃

Heating time

Room temperature ~400℃ 30min

Room temperature ~600℃   
 50min

Cooling time

400℃~Room temperature 

     (Equipped with rapid cooling device takes only 50min)

Stability time

8min

Display resolution

0.01℃

Power supply

220VAC/50HZ/800W

Communication Interface

RS-232

Sensor type

AA Level Pt100

Heating furnace size

280 (L)×270 (W) ×190 (H)(mm)

Controller size(mm)

218 (L) ×175 (W) ×145 (H)

Weight

3.5kg

Storage conditions

?20℃~60℃ (-4℉~140℉)

     5%RH-80%RH(No condensation)


Surface Sensor Calibrator - Surface Temperature Probe Calibration System
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
This article is read 4206 Times
Previous:Multifunction Process Calibrator | DTE-35G next:Custom temperature and humidity verification/calibration chamber

Contact us

DEARTO QR codeHeadquarter Address Tai'an High-tech Development Zone,  Shandong Province 
Telephon :+86 0538-5089056 +86 13605384645  +86 13953856217 
Fax : +86 0538-5059718 
After sale:+86 0538-5050959 +86 13605384645 
Website : kqcgg.cn 
E-mail : tadt17@dearto.cn /  tadtzdh@dearto.cn

Copyright? Tai'an Detu Automation Instrument Co., Ltd.             Lu ICP 14024757-1           Business license inquiry